How HPLC works: reading the peaks

The percentage on a peptide lab report measures area on a chart. Here is how the machine draws it, how to read it and what it misses.

The short answer

HPLC purity is the share of a chromatogram's total peak area that sits under the main peak.6 A chromatogram is the chart an HPLC machine draws while a UV detector watches a sample's substances leave its column, one after another.3

So the percentage says how much of what the detector saw is the main compound. Which compound that is, and how many milligrams the vial holds, take other tests. For peptides the detector usually reads at 210–220 nm, where the bonds joining amino acids absorb strongly.1

  • 210–220 nmwhere the detector reads peptides
  • 4working parts: pump, valve, column, detector
  • 3–10 µmsilica grains packed in the column

Chrysalis Labs Research Team8 min readUpdated

Strictly for in-vitro laboratory research. Not for human or veterinary use.

HPLC at a glance

Full name
High-performance liquid chromatography
What it draws
A chromatogram: detector signal against time
The purity figure
Main peak area ÷ total peak area × 100
Detector for peptides
UV light at 210–220 nm
Usual mode for peptides
Reversed-phase, on C8 or C18 columns
The column
A steel tube 30–300 mm long and 2.1–4.6 mm wide, packed with silica grains 3–10 µm across
Pressure
Several hundred atmospheres

Method facts from a peptide HPLC chapter,1 an analytical chemistry textbook2,3 and a study of synthetic peptides.6

What does HPLC stand for, and what is inside the machine?

High-performance liquid chromatography. Behind the panels sit four working parts: a pump, a valve that loads the sample, a column and a detector.2

The column does the sorting. It is a stainless-steel tube packed with porous silica grains 3 to 10 micrometres across.2 Pushing liquid through a bed that fine takes several hundred atmospheres, which is the pump's job.2

For peptides, the usual mode is reversed-phase.1 The grains carry carbon chains 8 or 18 atoms long (C8 or C18), which make their surface oily.2,1 The liquid is water with a rising share of acetonitrile, a lab solvent, plus a fraction of a percent of trifluoroacetic acid (TFA).1

Inside an HPLC system

  1. PumpDrives the liquid at several hundred atmospheres
  2. Sample valveLoads the sample into the flow
  3. ColumnHolds each substance back for its own time
  4. UV detectorReads the liquid at 210–220 nm
  5. ChartPlots signal against time
A drawing, not to scale.

Why does each substance come out at its own time?

Because each one grips the column differently. The column is oily and the liquid watery, so water-loving molecules leave first and water-shy ones linger.2 Peptides come off in order of how water-shy (hydrophobic) the whole chain is.1

The gradient. A run starts mostly as water. Minute by minute the pump raises the share of acetonitrile, and the liquid pulls oily molecules loose more easily.2,1 Each peptide lets go at its own point on that climb. The moment its peak tops out is its retention time.3

How fine is the sorting? In one study, peptides that differed by as little as one carbon atom came out as separate peaks once the method was tuned.1 Glycine for alanine is exactly that kind of swap, and it is the whole difference between GHK-Cu and AHK-Cu.

Why it matters. Many impurities are close relatives. Synthesis can leave chains with an amino acid missing or extra, mirror-image forms of an amino acid, and oxidised side chains.4 Being so alike, they tend to come out near the main peak.1

How do you read a chromatogram?

Left to right, like a timeline. The bottom axis is minutes since the sample went in. The side axis is the detector's signal: how strongly the liquid leaving the column absorbs UV light at that moment.3,2

A flat stretch, the baseline, means nothing is passing the detector. Each hill is a substance on its way out, described by three things:

  • Retention time: when it tops out, which places it on the timeline.3
  • Height: how strong the signal got.
  • Area: the space under the hill. For one substance it grows in step with the amount, so area is the usual measure.5

Example trace, read at 214 nm

  1. 1Peak 1, a small peak at 9.8 minutes.
  2. 2Peak 2, a small peak at 12.1 minutes, just before the main peak.
  3. Main peakpeak 3, 12.6 minPeak 3, the main peak, at 12.6 minutes.
  4. 4Peak 4, a small peak at 13.4 minutes, just after the main peak.
  5. 5Peak 5, a small peak at 16.0 minutes.
Example only: made-up peaks over a 20-minute run, not from a real report or product. Peaks 1 to 5 sit at 9.8, 12.1, 12.6, 13.4 and 16.0 minutes; the small ones are drawn taller than scale so they can be seen.

Each substance draws its own peak,3 so a single-compound vial should show one main peak and a blend one per compound. BPC-157 vs TB-500, and the blend vial shows two.

How is HPLC purity worked out?

By area percent. Divide the main peak's area by the total area of all counted peaks, then multiply by 100.6

The example, peak by peak: area and share

  1. Peak 1 · 9.8 min15 · 1.5%
  2. Peak 2 · 12.1 min35 · 3.5%
  3. Main peak · 12.6 min920 · 92.0%
  4. Peak 4 · 13.4 min20 · 2.0%
  5. Peak 5 · 16.0 min10 · 1.0%
Example only: made-up areas, not from a real report or product. Total area 1,000; the main peak's share is 920 ÷ 1,000 × 100.

In the example: 920 ÷ 1,000 × 100 = 92.0%. The two largest minor peaks sit either side of the main one, at 12.1 and 13.4 minutes, which is where close relatives of the main chain tend to land.

Why is HPLC for peptides read at 210–220 nm?

Every peptide is held together by bonds between its amino acids, and those bonds absorb strongly in that far-UV band,1 so nearly any peptide-like impurity shows up.

Tyrosine, phenylalanine and tryptophan also absorb around 250–290 nm, but not every peptide contains them.1 The wavelength shapes every signal,2 so a figure read only up there can miss whole peaks. Check the wavelength line on the report.

What does HPLC purity not tell you?

Four limits, true of any HPLC purity figure.

Which molecule the main peak is. A retention time that matches a known standard suggests a substance is there. It does not prove it.5 Identity usually takes mass spectrometry: molecules pick up an electric charge and are sorted by mass divided by charge.7 Coupled to HPLC, it reads what is inside a peak.7,2 For a worked case, see the five numbers in What is BPC-157?

What hides under the peak. Two substances leaving the column at nearly the same moment merge into one hill,3 and the area counts both as the main compound. A better-tuned method can pull them apart, and a mass spectrometer reading the peak can tell them apart if their masses differ.1,2,7

Unequal brightness. Area percent treats every substance as equally visible to the detector. At 214 nm each peptide bond absorbs a little, and some amino acids far more: tryptophan about 30 times a peptide bond, phenylalanine, tyrosine and histidine about 6 times.8 So equal amounts can draw unequal peaks. Exact amounts need a calibration standard of that same substance.2

What the share leaves out. Most peptides pull water from the air,9 and TFA, used to cut synthetic peptides free and to purify them, leaves many as trifluoroacetate salts.10 Water and the counter-ions in those salts get their own tests, such as Karl Fischer titration and ion chromatography.9,10

A 2019 multi-lab study shows the sums. To find how much of one peptide reference material was the peptide itself, the labs took its HPLC impurities, then subtracted acetic acid (about 7.5–8% of the weight on its own), leftover solvents and inorganic residue, measuring water separately.9

Three tests, three questions about one sample. Methods from sources 1, 2, 6, 7, 9 and 10.
FactPurity (HPLC)Identity (mass spec)Content
QuestionHow much of what the detector saw is the main peak?Is the main peak the expected molecule?How much of the powder is peptide?
HowTime on a column, UV at 210–220 nmMass divided by chargeAgainst a reference; water and counter-ions tested apart
You seeA trace and a % of total areaMass found against mass expectedmg, or a % of the powder's weight
Blind spotHidden peaks; unequal brightness; no UV signal, no peakAmounts, without an internal standardRelated impurities can count as peptide

Purity and identity: two different questions

What should the HPLC part of a lab report show?

Enough to redo the sum yourself.

The HPLC part of a report, line by line

  • Name and batchA compound name and batch number that match the vial's label.
  • Date and labThe test date and the lab that ran it.
  • MethodColumn type, solvent gradient and detection wavelength, usually 210–220 nm for peptides.
  • The chromatogramThe chart itself, both axes labelled, the whole run shown.
  • A peak tableRetention time, area and area % for every counted peak, adding up to 100.
  • Identity on the same batchUsually mass spec, with the mass found next to the mass expected.

6 lines to look for

What lets you check the sum yourself, on any report.

One sample, one date. A report describes the sample that was run, from one batch, on that day. Peptides can break down over time, and breakdown products are among the impurities a separation is there to catch.4 More in Do sealed peptide vials expire?

To verify a report, ask the lab that issued it. How to read a certificate of analysis

Chrysalis Labs is lab-verified.

In short

  • HPLC purity is the main peak's share of all peak area on a chromatogram: main peak area ÷ total peak area × 100.
  • Peptides are usually run in reversed-phase mode and read by UV at 210–220 nm, where peptide bonds absorb.
  • The percentage does not name the molecule (mass spec does) or weigh the peptide in a vial (content tests do).

Share this guide

Quick answers

What does HPLC stand for?

High-performance liquid chromatography. A pump drives a liquid sample through a packed column, which holds each substance back for its own time, and a detector draws each one as a peak on a chromatogram.

What does HPLC purity mean?

The main peak's share of all peak area on a chromatogram, read by a UV detector at one wavelength, usually 210–220 nm for peptides. It is worked out as main peak area ÷ total peak area × 100.

What does “HPLC tested” mean?

That a sample from a batch was run through HPLC and its peaks measured, giving a share of peak area. Identity and the amount of peptide need their own tests: look for a mass spec line and the batch number. Strictly for in-vitro laboratory research. Not for human or veterinary use.

Why is HPLC for peptides read at 210–220 nm?

The bonds joining amino acids absorb strongly there, so almost every peptide and peptide-like impurity gives a signal. Only a few amino acids, such as tyrosine and tryptophan, also absorb at longer wavelengths.

Does HPLC purity prove a peptide is the right one?

No. A retention time places a peak without naming it, and two substances can leave the column together. Identity is usually checked by mass spectrometry, which sorts molecules by mass divided by charge.

Is HPLC purity the same as the amount of peptide in a vial?

No. Purity is a share of peak area; the powder also holds water and counter-ions such as acetate, measured by other tests. In one 2019 multi-lab study, acetic acid alone was about 7.5–8% of a peptide reference material's weight.

  1. Mant et al. · Methods in Molecular Biology · 2007Reversed-phase is the most widely used HPLC mode for peptides; peptides leave in order of overall hydrophobicity; acetonitrile gradients with TFA; peptide bonds absorb strongly in the far UV, so detection is generally at 210–220 nm, while tyrosine, phenylalanine and tryptophan absorb around 250–290 nm. Peptides one carbon atom apart (glycine, alanine, valine, isoleucine swaps) were all resolved once the method was tuned. Closely related synthetic impurities run near the product.
  2. Harvey · Analytical Chemistry 2.1, section 12.5 · LibreTextsPump, sample loop, column and detector. Analytical columns 30–300 mm long and 2.1–4.6 mm wide, packed with 3–10 µm silica; pressures of several hundred atmospheres. C8 and C18 phases; more polar solutes leave first; gradients move from more to less polar. The wavelength affects each analyte's signal. A mass spectrometer detector gives structural information, and HPLC–MS needs an internal standard to measure amounts. Calibration uses standards.
  3. Harvey · Analytical Chemistry 2.1, section 12.2 · LibreTextsA chromatogram plots the detector's response against elution time, with a peak for each solute. Retention time runs from the sample going in to the top of its peak. Poorly resolved peaks overlap.
  4. D'Hondt et al. · Journal of Pharmaceutical and Biomedical Analysis · 2014Impurities of synthetic peptides: amino acids missing or inserted, diastereomers from racemisation, oxidised side chains, counter-ions such as trifluoroacetate, and breakdown products.
  5. Instrumental Analysis, section 10.1.7 · Sewanee: The University of the South · LibreTextsA peak's area is proportional to the amount of the substance. A matching retention time suggests a substance is present but is not definitive evidence.
  6. Verbeke et al. · Journal of Pharmaceutical Analysis · 2015Synthetic peptides checked by HPLC with UV detection: purity calculated as the UV area of the most abundant peak as a percentage of all peak areas (area normalisation), read at 210–215 nm.
  7. Banerjee and Mazumdar · International Journal of Analytical Chemistry · 2012Electrospray charges molecules, generally by adding hydrogen ions; the analyser sorts ions by mass to charge ratio; HPLC coupled with electrospray mass spec handles complex samples.
  8. Kuipers and Gruppen · Journal of Agricultural and Food Chemistry · 2007At 214 nm the peptide bond absorbs (923 M⁻¹ cm⁻¹); tryptophan absorbs about 30 times more, phenylalanine, tyrosine and histidine about 6 times more, so peptides differ in signal per amount.
  9. Li et al. · Journal of Pharmaceutical and Biomedical Analysis · 2019A study across 14 labs of one peptide reference material: most peptides are very hygroscopic; water measured by Karl Fischer titration; acetic acid 7.52–7.97% of the weight; peptide share found by mass balance from HPLC impurities, acetic acid, residual solvents and inorganic residue.
  10. Sikora et al. · Pharmaceuticals · 2020TFA is used to cleave and purify synthetic peptides, which are then obtained as trifluoroacetate salts; TFA is a routine HPLC additive. Peptide content, water and counter-ions should each be determined; ion chromatography is the first choice for counter-ions.

Our own pages: Purity and identity · Delivery zones and charges · FAQ · the catalogue

See our vials and pens

Chrysalis Labs is lab-verified and delivers to all seven emirates. Cash on delivery on Vials and Pens, as of .

Strictly for in-vitro laboratory research. Not for human or veterinary use.

Chrysalis Labs Research Team

More in Lab reports & labels

All in Lab reports & labels

Back to the Blog